DSO138 oscilloscope connection guide
DSO138 oscilloscope DSO138 13803K / 13804K wiring guide with power, signal assignments, and bench checks.
DSO138 13803K / 13804K
7 contacts Current packagePower the oscilloscope
RequiredUse a regulated 9 V DC adapter with a center-positive 5.5 × 2.1 mm plug and a current rating above 200 mA.
- Adapter positive
- DC_IN (J10 center)
- Adapter +
- Adapter return
- DC_RETURN (J10 sleeve)
- Adapter −
Bench checks
- Verify the polarity and the unloaded adapter voltage before inserting the plug. The kit accepts 8–12 V DC.
- Power on with the BNC probe disconnected and set coupling to GND. Confirm that the display starts and shows a steady zero trace.
JYE Tech DN138-01v05, Operation and Probe Calibration on p. 3 and schematic on p. 4.
Connection scenarios
Measure a 1.5 V cell
Set DC coupling and start at 1 V/div with a ×1 probe. Connect a single loose 1.5 V cell directly to the probe clips.
- Cell positive
- AIN (J1 center)
- Cell +
- Cell return
- AGND (J1 shell)
- Cell −
Bench checks
- Record the zero position with coupling set to GND, then return the switch to DC.
- The trace moves upward from zero by the cell voltage divided by the volts-per-division setting. Compare the displayed reading with a multimeter if you have one.
JYE Tech DN138-01v05, Operation and Probe Calibration on p. 3 and schematic on p. 4.
DSO138 13803K / 13804K wiring map
J1 BNC input
J10 DC barrel jack
J9 alternate DC header
J2 test output
J9-1 · DC_IN
Positive contact of the alternate PCB power header.
Models and documents
This map covers original JYE Tech kits 13803K and 13804K. The optional J4 Mini-USB footprint is listed without a user function. Use J10 or J9 for power.
- DSO138 13803K / 13804K
- JYE Tech kit connectors for the BNC input, DC supply and test signal.
- JYE Tech DSO138 user manual· JYE Tech DN138-01v05, Operation and Probe Calibration on p. 3 and schematic on p. 4.
- JYE Tech DSO138 product documentation· Kit revisions, input characteristics and DC connector dimensions.